Review and approve an uncertainty budget
The author submits a complete draft for review. A different authorised member checks the range, exact unit, component classification, input mode, source type, distributions, divisors, sensitivity coefficients, degrees of freedom, fixed coverage policy, and effective dates.
- Open Calibration, then Uncertainty budgets.
- Open the budget and inspect the draft revision and every ordered component.
- If you are the author, select Submit for review using the current revision version.
- Have a different authorised member reopen the budget and select Approve revision.
Approval freezes the revision content and integrity hash. A version conflict means another member changed the lifecycle first; reload before deciding what to do.
If a submitted revision is unsuitable, an independent reviewer can Reject for correction with a reason. The author can also Withdraw for correction. The reviewed content is retained; use Create replacement draft to correct it and submit the replacement for a fresh review. Approval of the replacement supersedes the previous approved revision.
The author or submitter cannot approve their own revision. Approval of a newer revision does not alter calculations already frozen against an older revision.
Next: Assign an uncertainty budget.